Fluke-983 파티클 카운터 |
|
|
|
|
Specifications |
Channels |
|
0.3, 0.5, 1.0, 2.0, 5.0, 10.0 µm |
|
Flow Rate |
|
0.1 cfm (2.83 L/min) controlled by internal pump |
|
Count Modes |
|
Concentration, totalize, audio |
|
Counting Efficiency |
|
50% @0.3 µm; 100% for particles > 0.45 µm (per JIS B9921:1997) |
|
Zero Count |
|
1 count/5 minute (JIS B9921:1997) |
|
Coincidence Loss |
|
5% at 2,000,000 particles per ft3 |
|
Relative Humidity |
|
± 7%, 20% to 90% non-condensing |
|
Temperature |
|
± 3 ˚C, 10 ˚C to 40 ˚C (50 ˚F to 104 ˚F) |
|
Data Storage |
|
5000 sample records (rotating buffer) of date, time, counts, relative humidity, temperature, sample volumes, alarms, label |
|
Alarms |
|
Counts, low battery, sensor fail |
|
Delay Time |
|
Sample Inlet |
|
Interface |
|
RS-232 and RS-485 via RJ-45 |
|
Environmental Specifications |
Operating: |
|
10 ˚C to 40 ˚C (50 ˚F to 104 ˚F), 20% to 90% relative humidity, non-condensing |
|
Storage: |
|
-10 ˚C to 50 ˚C (14 ˚F to 122 ˚F), up to 90% relative humidity, non-condensing |
|
Mechanical & General Specifications |
Size |
|
11.43 cm x 20.96 cm x 5.72 cm (4.5 in x 8.25 in x 2.25 in) |
|
Dimensions (W x L x D) |
|
Weight |
|
Power |
|
AC adapter, 12 V DC at 2.5A, 90 to 250 V AC, 50 to 60 Hz |
|
Rechargeable Battery |
|
NiMH, 4.8 V at 4.5 Ah; replaceable |
|
Battery Life |
|
Recharge Time |
|
Calibration |
|
PSL particles in air (NIST traceable) |
|
기기를 포함한 품목:
· 교정 성적서 (NIST)
· Windows와 호환되어 다운로드 되어지는 소프트웨어
· DB9 to RS-232 아답터와 케이블
· Isokinetic 브로브
· 영점 카운터 필터
· 고순도 튜브
· 1/8 인치 가시모양 호스 어답터
· 파워 서플라이
· 매뉴얼
· 하드 케이스
|
|
|
|
|
|